Oxygen signal: Difference between revisions

From Bioblast
(Created page with "{{MitoPedia |description=The '''oxygen signal''' of the O2k is transmitted from the electrochemical polarographic oxygen sensor (OroboPOS) for each of the two chambers to...")
ย 
No edit summary
Line 20: Line 20:
* Confirm and locate the problem by performing a [[Sensor test]] and following [[Locating a problem]].
* Confirm and locate the problem by performing a [[Sensor test]] and following [[Locating a problem]].
: Follow the standard instructions for [[MiPNet19.18B POS-Service|oxygen sensor service]].
: Follow the standard instructions for [[MiPNet19.18B POS-Service|oxygen sensor service]].
=== Oxygen signal does not respond ==
* The oxygen signal remains at or close to zero even at high oxygen levels.
* The signal at air saturation is abnormally low.
:# Check the [[gain]] setting for the affected oxygen channel.
::::* Response 1: The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::* Response 2: The gain was correct: Locate the problem using the procedure described in [[Locating a problem]] and [[Sensor test]].

Revision as of 23:01, 6 February 2016


high-resolution terminology - matching measurements at high-resolution


Oxygen signal

Description

[[Description::The oxygen signal of the O2k is transmitted from the electrochemical polarographic oxygen sensor (OroboPOS) for each of the two chambers to DatLab. The primary signal is a current mAmp which is converted into a voltage [V], and calibrated in SI units for amount of substrance concentration [ยตmol.dm-3 or ยตM].]]


Reference: [[Info::MiPNet19.18D O2k-Calibration, MiPNet06.03 POS-Calibration-SOP]]



MitoPedia methods: Respirometry, DatLab 

Template:MitoPedia SUIT



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Stability of the oxygen signal

The stability of the oxygen signal of the O2k is evaluated by the (negative) uncorrected slope over time. After smoothing, the slope of a stable sensor should be less than +- 1 pmol.s-1.ml-1 in a 2-ml chamber.

  • Instability of the oxygen signal
  • There is drift of the oxygen signal over short or long periods of time.
  • The signal at air saturation is abnormally high for a given gain setting.
  • The zero current is high (>2.5%) and may show drift.
Follow the standard instructions for oxygen sensor service.

= Oxygen signal does not respond

  • The oxygen signal remains at or close to zero even at high oxygen levels.
  • The signal at air saturation is abnormally low.
  1. Check the gain setting for the affected oxygen channel.
  • Response 1: The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
  • Response 2: The gain was correct: Locate the problem using the procedure described in Locating a problem and Sensor test.
Cookies help us deliver our services. By using our services, you agree to our use of cookies.