MiPNet19.18D O2k-Series G and DatLab 6: Calibration: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 20: Line 20:
procedure is described for O2k high-resolution respirometry with the
procedure is described for O2k high-resolution respirometry with the
automatic calibration routine of DatLab.
automatic calibration routine of DatLab.
:>> O2k-Manual:[http://www.oroboros.at/?o2k-manualΒ  Overall contents]
:>> O2k-Manual: [http://www.oroboros.at/?o2k-manualΒ  Overall contents]
:>> Product: [http://www.oroboros.at/?oxygraph OROBOROS Oxygraph-2k], [[O2k-Catalogue_OROBOROS| O2k-Catalogue]]
:>> Product: [http://www.oroboros.at/?oxygraph OROBOROS Oxygraph-2k], [[O2k-Catalogue_OROBOROS| O2k-Catalogue]]
|keywords=HRR, DatLab
|keywords=HRR, DatLab

Revision as of 10:14, 17 October 2013

Publications in the MiPMap
Gnaiger E (2013) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08.

Β» [[Has info::Open Access-MiPNet12.08.pdf]], [[Has info::Versions

O2k-Manual contents

]]

OROBOROS (2013) Mitochondr Physiol Network

Abstract: [[has abstract::O2k-Manual Chapter D

Accurate calibration of the oxygen sensor depends on (1) calibration solutions prepared at known oxygen partial pressures, as achieved in the Oxygraph-2k (O2k) at defined temperature, continuously recorded total gas pressure (barometric pressure), and thermodynamic equilibrium between the gas and aqueous phase; (2) high stability of the signal of the polarographic oxygen sensor (POS), tested for sufficiently long periods of time; (3) linearity of signal output with oxygen pressure, achieved with the POS in the range between oxygen saturation and zero oxygen pressure; and (4) application of accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration (MiPNet06.03). The standard oxygen calibration procedure is described for O2k high-resolution respirometry with the automatic calibration routine of DatLab.

>> O2k-Manual: Overall contents
>> Product: OROBOROS Oxygraph-2k, O2k-Catalogue]]

β€’ Keywords: HRR, DatLab

β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual, Protocol 

O2k-SOP, DatLab 


Further information

>> MiPNet12.08
>> MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) are integrated into MiPNet12.08.
  • Version 6: 2012-10-24
  • Version 1: 2007