O2k-Barometric Pressure Transducer: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 17: | Line 17: | ||
== Troubleshooting == | == Troubleshooting == | ||
:::: If the barometric pressure is displayed as 0 in DatLab, switch of the O2k and unplug it from the socket for at least 15 minutes. In most cases the problem is resolved after connecting the O2k to the socket and switching it on again. If the problem persists or happens again, please contact our [https://www.oroboros.at/index.php/o2k-technical-support/| technical support]. | :::: If the barometric pressure is displayed as 0 kPa in DatLab (O2 calibration window), switch of the O2k and unplug it from the socket for at least 15 minutes. In most cases the problem is resolved after connecting the O2k to the socket and switching it on again. If the problem persists or happens again, please contact our [https://www.oroboros.at/index.php/o2k-technical-support/| technical support]. | ||
{{MitoPedia O2k and high-resolution respirometry | {{MitoPedia O2k and high-resolution respirometry | ||
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry | |mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry | ||
}} | }} |
Revision as of 14:18, 24 August 2023
O2k-Barometric Pressure Transducer |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years.
Reference: O2k-Core
- SI unit: Pa = Jยทm-3
- 1 atm = 760 mmHg = 101.325 kPa
Accuracy
- Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
- 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
- Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa
Troubleshooting
- If the barometric pressure is displayed as 0 kPa in DatLab (O2 calibration window), switch of the O2k and unplug it from the socket for at least 15 minutes. In most cases the problem is resolved after connecting the O2k to the socket and switching it on again. If the problem persists or happens again, please contact our technical support.
MitoPedia O2k and high-resolution respirometry:
O2k hardware,
O2k-Open Support,
O2k-Respirometry,
O2k-FluoRespirometry