Difference between revisions of "Open chamber"
From Bioblast
m |
|||
Line 1: | Line 1: | ||
{{Technical service}} | |||
The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 [[flux at open chamber]]. | The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 [[flux at open chamber]]. | ||
To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a [[stopper spacer]]. | To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a [[stopper spacer]]. | ||
Line 10: | Line 12: | ||
{{#set:Technical service=Chamber| Technical service=POS calibration}} | {{#set:Technical service=Chamber| Technical service=POS calibration}} | ||
[[Category:Technical service]] | |||
__SHOWFACTBOX__ | __SHOWFACTBOX__ | ||
Revision as of 16:57, 13 February 2015
The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 flux at open chamber. To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a stopper spacer. This procedure is described in detail in the chapter "Air calibration" in [MiPNet12.08].
See also: