'''The O<sub>2</sub> sensor test is described as an [[OroboPOS]]''' standard operating procedure: '''Polarographic oxygen sensors: calibration, accuracy and quality control SOP.''' ยป[[MiPNet06.03 POS-Calibration-SOP]]ยซ
ย 
== When should an O2-sensor test be performed? ==
* After switching on the O2k, every day: air calibration and stirrer test.
* Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen.
* After application of a new membrane and O2-sensor service: In some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time), when the O2k remains switched on over night (O2k-Chambers filled with 70% EtOH).
* During troubleshooting, particularly when [[switching components]] between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).
ย 
== The main criteria for sensor performance ==
ย 
* [[Raw signal at air saturation]]
* [[Zero current |Raw signal at zero oxygen]]
* [[Signal stability]]: drift and noise
* [[Response time of the POS]]
ย 
ย 
ย 
[[Category:Technical service]]
ย 
{{#set:Technical service=O2 sensor|Technical service=POS connector|Technical service=O2 signal|Technical service=Main unit|Technical service=Performance test}} ย 
__SHOWFACTBOX__

Revision as of 20:39, 19 May 2015


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



ยป MiPNet06.03 POS-Calibration-SOP
Cookies help us deliver our services. By using our services, you agree to our use of cookies.