==Problems==
#REDIRECT [[Oxygen signal]]
*The oxygen signal remains at or close to zero even at high oxygen levels.
*The signal at air saturation is abnormally low.
 
==Test and Solutions==
 
'''The oxygen signal is very low at air saturation and less responsive than expected.'''
:'''Solution 1:''' Check the [[gain]] setting for the affected oxygen channel.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a Problem]] and [[Sensor test]].
 
 
{{#set: Technical service= Sensor|Technical service= O2 signal}}
 