Oxygen signal not responding: Difference between revisions

From Bioblast
mNo edit summary
mNo edit summary
Line 8: Line 8:
:'''Solution 1:''' Check the [[gain]] setting for the affected oxygen channel.
:'''Solution 1:''' Check the [[gain]] setting for the affected oxygen channel.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a Problem]] and [[Sensor test]]
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a Problem]] and [[Sensor test]].





Revision as of 15:32, 14 September 2010

Problems

  • The oxygen signal remains at or close to zero even at high oxygen levels.
  • The signal at air saturation is abnormally low.

Test and Solutions

The oxygen signal is very low at air saturation and less responsive than expected.

Solution 1: Check the gain setting for the affected oxygen channel.
  • The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
  • The gain was correct: Locate the problem using the procedure described in Locating a Problem and Sensor test.



Template:Technical service


Cookies help us deliver our services. By using our services, you agree to our use of cookies.