POS (technical service): Difference between revisions

From Bioblast
mNo edit summary
(Redirected page to OroboPOS)
 
(4 intermediate revisions by the same user not shown)
Line 1: Line 1:
As the heart of the OROBOROS Oxygraph the [[Polarographic oxygen sensor]] is of course the prime suspect when any kind of troubles are encountered. However, frequently problems can be traced either to inadequate software settings or to other hardware components (like the [[POS connector (technical service) |POS connector]]).
#REDIRECT [[OroboPOS]]
Therefore it is important to confirm the problem doing a [[Sensor test]] and to [[Locating a problem |localize the problem]].
 
==Possible Problems==
===Problems that require a full Sensor test and Localization of the Problem===
*unstable O2 signal
*unstable O2 flux
*high O2 signal at air saturation
*high O2 signal at zero oxygen
 
===Special Cases: no further localization necessary ===
* [[Slow responding sensor]], if this has been confirmed by a [[Stirrer test]]. Note: The stirrer test is part of the sensor test. It is therefore recommended to do a full [[Sensor test]] whenever problems are encountered.
 
==Performing the Sensor Test==
 
Confirm the problem by doing a [[Sensor test]] in the absence of biological material
 
===Option 1.A: No problem was visible in the [[Sensor Test]]===
check the following DatLab settings
* [[Gain (O2 channel)]]
* [[POS calibration]]
* Scaling
 
If the problem is not visible when observing the [[Raw signal]] but is visible when observing the calibrated oxygen signal, then there is probable a problem with the [[POS calibration]].
 
===Option 1.B: The problem was visible in the [[Sensor Test]]===
* proceed to Localizing the Problem
 
==Localizing the Problem==
 
[[ Locating a problem|Locate the Problem ]]
 
===Option 2.A: The problem could be localized on the [[POS]]===
 
'''Solutions 2.A'''
Follow the instructions for oxygen sensor service [http://www.oroboros.at/index.php?id=o2k-pos-service [MiPNet08.04]]. Use this opportunity to apply [[contact oil]] to the gold pin and thread connecting [[POS]] and [[sensor connector]]. After a sensor service or after applying anew membrane allow the oxygraph to run over night to achieve a stable signal before repeating the [[sensor test]].
If all service precautions fail to solve the problem, the [[POS]] has to be sent to us for further service (a recovery can not be guaranteed), or a new sensor is to be applied.
 
===Option 2.B: The problem was not localized on the [[POS]]===
'''Solutions 2.B'''
If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.
 
 
 
 
 
 
{{#set: Technical service= Sensor|Technical service= O2 signal|Technical service= Sensor service| Technical service= O2 flux}}
 
{{Technical service}}
 
 
__SHOWFACTBOX__

Latest revision as of 14:54, 14 February 2012