DatLab converts the [[raw signal]] to oxygen concentrationsΒ  by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
#REDIRECT [[MiPNet06.03 POS-calibration-SOP]]
Please see [[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]] Β 
Β 
'''Dynamic POS calibration''' = Calibration of the sensor response time.
'''Static POS calibration''' Two-point calibration of the polarographic oxygen sensor.