Difference between revisions of "Polarization voltage"
From Bioblast
(added technical support banner) Β |
|||
(23 intermediate revisions by 6 users not shown) | |||
Line 1: | Line 1: | ||
{{Technical support}} | |||
The default value is 800 mV. | {{MitoPedia without banner | ||
Β | |abbr=''U'' | ||
Β | |description=A '''polarization voltage''' of 600 mV to 800 mV is applied between anode and cathode of the [[polarographic oxygen sensor]], resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage. | ||
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]] | |||
Β | }} | ||
Β | {{MitoPedia O2k and high-resolution respirometry | ||
|mitopedia O2k and high-resolution respirometry=O2k hardware, DatLab | |||
}} | |||
{{Technical support integrated}} | |||
== DatLab == | |||
:::: The polarization voltage is set in [[DatLab]] in the [[O2k control]] window [F7] ([[MiPNet22.11 O2k-FluoRespirometer manual]]). The default value is 800 mV. |
Latest revision as of 09:20, 5 April 2022
Polarization voltage |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
Abbreviation: U
Reference: MiPNet22.11 O2k-FluoRespirometer manual
MitoPedia O2k and high-resolution respirometry:
O2k hardware,
DatLab
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
DatLab
- The polarization voltage is set in DatLab in the O2k control window [F7] (MiPNet22.11 O2k-FluoRespirometer manual). The default value is 800 mV.