Difference between revisions of "Stirrer-Bar\white PVDF\15x6 mm"
From Bioblast
Line 10: | Line 10: | ||
* For dispatch of the [[O2k-Core]], the Stirrer-Bars are shipped in the [[O2k-Service Box]]. | * For dispatch of the [[O2k-Core]], the Stirrer-Bars are shipped in the [[O2k-Service Box]]. | ||
A "jumping" stirring bar problem is frequently caused by small particles sticking to the glass chamber, not the stirring bar itself | A "jumping" stirring bar problem is frequently caused by small particles sticking to the glass chamber, not the stirring bar itself. >> [[MiPNet19.03 O2k-cleaning and ISS]]. |
Revision as of 16:56, 14 April 2014
Stirrer-Bar\white PVDF\15x6 mm
Description | Stirrer-Bar\white PVDF\15x6 mm, operated in the O2k-Chamber at constant stirring speed (standard is 750 rpm, or 12.5 Hz), to provide optimum mixing of the sample in the aqueous medium and ensure a stable signal of the polarographic oxygen sensor (OroboPOS) placed in a position of maximum current of the medium.
O2k-Core: 2 in O2k-Assembly Kit. |
---|---|
Product ID | 23210-01 |
Type | O2k, O2k-Assembly Kit |
Link | MiPNet06.01 O2k-Overview |
Image |
- For dispatch of the O2k-Core, the Stirrer-Bars are shipped in the O2k-Service Box.
A "jumping" stirring bar problem is frequently caused by small particles sticking to the glass chamber, not the stirring bar itself. >> MiPNet19.03 O2k-cleaning and ISS.