Syringe\500 mm3 51/0.41 mm: Difference between revisions
From Bioblast
Beno Marija (talk | contribs) No edit summary |
Tindle Lisa (talk | contribs) No edit summary |
||
Line 4: | Line 4: | ||
|abbr= | |abbr= | ||
|description=[[File:Syringe 500 mm3 51 0.41 mm.JPG |right|180px]]Hamilton '''Syringe\500 mm<sup>3</sup> 51/0.41 mm''' for manual titrations, 500 mm<sup>3</sup> volume; fixed needle with rounded tip: 51 mm length, 0.41 mm inner diameter; for injections of suspensions of isolated mitochondria and filling of the [[Microsyringe\200 mm3\TIP2k]]. | |description=[[File:Syringe 500 mm3 51 0.41 mm.JPG |right|180px]]Hamilton '''Syringe\500 mm<sup>3</sup> 51/0.41 mm''' for manual titrations, 500 mm<sup>3</sup> volume; fixed needle with rounded tip: 51 mm length, 0.41 mm inner diameter; for injections of suspensions of isolated mitochondria and filling of the [[Microsyringe\200 mm3\TIP2k]]. | ||
|info=[ | |info=[https://www.oroboros.at/index.php/product/syringe500-mm3-51-0-41-mm/ '''Product details and purchase information'''] | ||
}} | }} | ||
Latest revision as of 11:09, 11 November 2020
Syringe\500 mm3 51/0.41 mm |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
Hamilton Syringe\500 mm3 51/0.41 mm for manual titrations, 500 mm3 volume; fixed needle with rounded tip: 51 mm length, 0.41 mm inner diameter; for injections of suspensions of isolated mitochondria and filling of the Microsyringe\200 mm3\TIP2k.
Reference: Product details and purchase information
Technical support
- Safety information: fixed needles are used with rounded tips, not usable for injection into tissues.
- Titration of suspensions of isolated mitochondria: larger inner diameter avoids mechanical damage of the mitochondria.
- Titration of large volumes of stock solutions into the O2k-Chamber (e.g. succinate at 50 mM final concentration).
- TIP2k-Module: gas bubble-free and gas-tight filling of the Microsyringe\200 mm3\TIP2k and the Microsyringe\500 mm3\TIP2k, with a minimum waste of the stock solution.
- For safe an clean handling, place the syringe in the one of the Syringe Racks.
- MiPNet19.14 - SOP for manual O2k-titrations with Hamilton microsyringes.
MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry