:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the βO2 slope negβ acceptable?''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the βO2 slope negβ acceptable?''' | ||
::::::: If you are not certain that the noise of the βO2 slope negβ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]] | ::::::: If you are not certain that the noise of the βO2 slope negβ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]] | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Did you know that you can use the [[TIP2k-Module|Titration-Injection microPump (TIP2k)]] to perform an instrumental O<sub>2</sub> background on the [[O2k-sV-Module|Small O2k-Chamber]] (0.5 mL)?''' | |||
::::::: By adjusting the dithionite concentration toΒ 2.5 mM, instrumental O2 background can be performed with the standard settings on the [[O2k-sV-Module|Small O2k-Chamber]] (0.5 mL).Β ''Further details in'' [[Dithionite]] and Β»[[MiPNet14.06 Instrumental O2 background]] |
Revision as of 12:52, 4 March 2020
For discussion
Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020
- Do you perform a stirrer test during oxygen calibration?
- In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in MitoPedia: Β»Oxygen calibration - DatLab
- Do you perform a stirrer test during oxygen calibration?
- Is the noise of the βO2 slope negβ acceptable?
- If you are not certain that the noise of the βO2 slope negβ is within a suitable range or you want to confirm that your experimental data will not be affected, see further details: MiPNet06.03 POS-calibration-SOP
- Is the noise of the βO2 slope negβ acceptable?
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the Small O2k-Chamber (0.5 mL)?
- By adjusting the dithionite concentration to 2.5 mM, instrumental O2 background can be performed with the standard settings on the Small O2k-Chamber (0.5 mL). Further details in Dithionite and Β»MiPNet14.06 Instrumental O2 background
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the Small O2k-Chamber (0.5 mL)?