| == For discussion ==
| |
| ''Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020''
| |
|
| |
| :::: [[File:MitoPedia.jpg|left|60px|MitoPedia]] As part of the [[NextGen-O2k|H2020 SME NextGen-O2k]] project we are developing a PhotoBiology module that will allow to study the effect of different light intensities on O<sub>2</sub> consumption and production: Β»'''[[PhotoBiology]]'''Β«
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you perform a stirrer test during oxygen calibration?'''
| |
| ::::::: In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. ''Further details in'' [[MitoPedia: DatLab |MitoPedia]]: Β»[[Oxygen calibration - DatLab]]
| |
|
| |
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the βO2 slope negβ acceptable?'''
| |
| ::::::: If you are not certain that the noise of the βO2 slope negβ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]]
| |
|
| |
|
| |