Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020
- Do you perform a stirrer test during oxygen calibration?
- In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in MitoPedia: Β»Oxygen calibration - DatLab
- Do you perform a stirrer test during oxygen calibration?
- Is the noise of the βO2 slope negβ acceptable?
- If you are not certain that the noise of the βO2 slope negβ is within a suitable range or you want to confirm that your experimental data will not be affected, see further details: MiPNet06.03 POS-calibration-SOP
- Is the noise of the βO2 slope negβ acceptable?
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the Small O2k-Chamber (0.5 mL)?
- By adjusting the dithionite concentration to 2.5 mM, instrumental O2 background can be performed with the standard settings on the Small O2k-Chamber (0.5 mL). Further details in Dithionite and Β»MiPNet14.06 Instrumental O2 background