Temperature plot empty: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|description='''Problem:''' Layout "01 Calibration Exp. Gr3-Temp" is selected, a third plot is displayed on the screen but it remains empty (no plot is shown). Newer versions of [[DatLab]] include pre-installed layouts which do not recognize some channel designations from older O2k series.
|description='''Problem:''' Layout "01 Calibration Exp. Gr3-Temp" is selected, a third plot is displayed on the screen but it remains empty (no plot is shown). Newer versions of [[DatLab]] include pre-installed layouts which do not recognize some channel designations from older O2k series.
|info=[[MiPNet19.18C DatLab Guide]]
|info=[[MiPNet19.18C DatLab guide]]
}}
}}
{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
|mitopedia O2k and high-resolution respirometry=DatLab
}}
}}
{{MitoPedia topics}}
{{Technical support integrated}}
{{Technical support integrated}}
 
== Action ==
The problem of an empty graph can be solved by selecting the approbate channels manually and then save the modified layout:
:::: The problem of an empty graph can be solved by selecting the approbate channels manually and then save the modified layout:
# Select layout "01 Calibration Exp. Gr3-Temp".
::::# Select layout "01 Calibration show Temp".
# Select the third, empty graph at the bottom by clicking on it.
::::# Select the third, empty graph at the bottom by clicking on it.
# In the menu go to "Graph" / "Select plots" and manually check "Block temp." and "Peltier Pow." for the Y1 and Y2 axis, respectively.
::::# In the menu go to "Graph" / "Select plots" and manually check "Block temp." and "Peltier pow." for the Y1 and Y2 axis, respectively.
# Select a line color and thickness for both plots.
::::# Select a line color and thickness for both plots.
# Save the graph layout, replacing "01 Calibration Exp. Gr3-Temp".
::::# Save the graph layout, replacing "01 Calibration show Temp".

Revision as of 09:50, 14 August 2016


high-resolution terminology - matching measurements at high-resolution


Temperature plot empty

Description

Problem: Layout "01 Calibration Exp. Gr3-Temp" is selected, a third plot is displayed on the screen but it remains empty (no plot is shown). Newer versions of DatLab include pre-installed layouts which do not recognize some channel designations from older O2k series.


Reference: MiPNet19.18C DatLab guide


MitoPedia O2k and high-resolution respirometry: DatLab 


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Action

The problem of an empty graph can be solved by selecting the approbate channels manually and then save the modified layout:
  1. Select layout "01 Calibration show Temp".
  2. Select the third, empty graph at the bottom by clicking on it.
  3. In the menu go to "Graph" / "Select plots" and manually check "Block temp." and "Peltier pow." for the Y1 and Y2 axis, respectively.
  4. Select a line color and thickness for both plots.
  5. Save the graph layout, replacing "01 Calibration show Temp".