Template Select fluorescence sensor

From Bioblast
Revision as of 09:35, 19 November 2021 by Komlodi Timea (talk | contribs) (โ†’โ€ŽSelect the Fluo-Sensors)
(diff) โ† Older revision | Latest revision (diff) | Newer revision โ†’ (diff)

Select the Fluo-Sensors

Switching between different excitation wavelengths and filters is achieved by simply exchanging the Fluo-Sensors. Two types of optical sensors are supplied with different LEDs for fluorescence excitation, and the effective spectra of the LEDs are modified by filters.
Select the Fluo-Sensor and filter set from the table on application-specific settings (here we should link to the respective section which is also a template). Each Fluo-Sensor is delivered with a mounted filter set:
  1. Fluorescence-Sensor Green and Smart Fluo-Sensor Green: 525 nm (dominant wavelength). The installed Filter-Cap is ready for H2O2 measurement with Amplexยฎ UltraRed or for measurements of mt-membrane potential with TMRM.
  2. Fluorescence-Sensor Blue and Smart Fluo-Sensor Blue: 465 nm (dominant wavelength). The installed Filter-Cap is ready for measurements of mt-membrane potential with safranin or Rhodamine 123. A different filter is used for measurement with Magnesium Greenโ„ข or Calcium Greenโ„ข.

PlayVideo.jpg Exchanging a filter set of the Smart Fluo-Sensor

Cookies help us deliver our services. By using our services, you agree to our use of cookies.