Volume-Calibration Ring: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 13: | Line 13: | ||
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring. | ::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring. | ||
[[Image:Volume calibration ring A+B aluminum for Stopper.jpg|right| | [[Image:Volume calibration ring A+B aluminum for Stopper.jpg|right|400px]] | ||
== Innovation == | == Innovation == | ||
:::: Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position. | :::: Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position. |
Revision as of 21:40, 23 October 2016
Volume-Calibration Ring
Description | Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers, white; attached to the stoppers. |
---|---|
Product ID | 2432A-01 or 2432B-01 |
Type | O2k, O2k-Assembly Kit, Stopper |
Link | MiPNet19.18A O2k-start |
Image |
Technical support
- Volume calibration of the O2k-Chamber
- Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
- Follow the instructions in the O2k-Manual: MiPNet19.18A O2k-start
- Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
- Volume calibration of the O2k-Chamber
Innovation
- Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position.