Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Zero calibration

From Bioblast
Revision as of 15:33, 21 September 2010 by Fasching Mario (talk | contribs) (created page)
(diff) ← Older revision | Latest revision (diff) | Newer revision β†’ (diff)

Zero calibration is together with air calibration one of the two steps of the POS calibration, see [MiPNet12.08]. It is done at closed chamber after all the oxygen has been removed by the addition of dithionite. A zero calibration is the last part of the sensor test and described in detail there. Unlike air calibration it is not necessary to perform a zero calibration each day.

A zero calibration should be performed

A zero calibration has to be performed

  • when a new or serviced POS is used for the first time
  • when a new membrane was applied to the sensor
  • when the gain is changed
  • when a new POS connector is used



Template:Technical service

... more about "Zero calibration"
POS calibration +