::::::: If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]] | ::::::: If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]] | ||
== Archive 2020 == | == Archive 2020 == |
Revision as of 15:21, 6 February 2020
For discussion
Here we discuss suggestions for the Oroboros Science calendar
- Do you perform a stirrer test during oxygen calibration?
- In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in MitoPedia: »Oxygen calibration - DatLab
- Do you perform a stirrer test during oxygen calibration?
- Is the noise of the ‘O2 slope neg’ acceptable?
- If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see further details: MiPNet06.03 POS-calibration-SOP
- Is the noise of the ‘O2 slope neg’ acceptable?
Archive 2020
Jan 2020
2020 Week 1 The Oroboros team enjoys holidays from 2019 Dec 24 to 2020 Jan 06
|
- Jan-04 Sa
- The Oroboros 2020 Science Agenda is launched as a signature of the Oroboros Ecosystem.
- One month ago
|
2020 Week 2
- Jan-07 Tu The Oroboros Office opens in the New Year 2020
- From the Austrian Bioenergetics Group to Mitochondrial Physiology and Bioenergetics DACH (D: Germany; A: Austria; CH: Switzerland)
- EBEC22 - the 22nd European Bioenergetics Conference will be held in Innsbruck (2022, Aug 22-27). We would like to invite several Austrian colleagues to join the EBEC22 National Scientific Committee. This provides an excellent opportunity to establish an Austrian Bioenergetics Group towards and beyond EBEC22. We hope that this topic will be discussed at the forthcoming 1st Myocardial Function Symposium 2020 Graz AT, and further at the MitoEAGLE Summit Obergurgl 2020. Please, contact the organizer of EBEC22 - [email protected]
- Jan-08 We
- Jan-09 Th
- Planning the MitoEAGLE Summit Obergurgl 2020
- Members of the Management Committee MitoEAGLE and Working Group Leaders: please submit your suggestions for MitoEAGLE Summit Obergurgl 2020:
- Suggested type and title of your personal presentation, according to the MoU and final MitoEAGLE output
- Suggested scientists (titles) for presentations, according to the MoU
- Members of the Management Committee MitoEAGLE and Working Group Leaders: please submit your suggestions for MitoEAGLE Summit Obergurgl 2020:
- Jan-10 Fr
- » 1st Myocardial Function Symposium 2020 Graz AT
- 14:45 Novel technologies for mitochondria and cell research (Gnaiger, Innsbruck, AUT)
- » 1st Myocardial Function Symposium 2020 Graz AT
- Jan-11 Sa - 1st Myocardial Function Symposium 2020 Graz AT
- Jan-12 So
- One month ago
|
2020 Week 3
- Jan-13 Mo
- Jan-14 Tu
- Jan-15 We
- Statistical evaluation on replica in HRR with permeabilized skeletal muscle fibers from the O2k-Network: AU Melbourne Stepto NK
- Jacques M, Kuang J, Bishop DJ, Yan X, Alvarez-Romero J, Munson F, Garnham A, Papadimitriou I, Voisin S, Eynon N (2019) Mitochondrial respiration variability and simulations in human skeletal muscle: The Gene SMART study. FASEB J [Epub ahead of print]. - »Bioblast link«
- Statistical evaluation on replica in HRR with permeabilized skeletal muscle fibers from the O2k-Network: AU Melbourne Stepto NK
- Jan-16 Th
- Oxygen flux - which SI units should be used?
- Did you ever think that units for body mass could be an issue? - More details: Body mass
- Jan-17 Fr
- HRR - high-resolution respirometry and oxygen kinetics 10 years ago: Cell respiration under hypoxia: facts and artefacts in mitochondrial oxygen kinetics. Adv Exp Med Biol 2010. - »Bioblast link«
- Jan-18 Sa - TRANSMIT Symposium 2020 Brussels BE
- Mitochondrial respiration is oxygen-dependent in human cardiac homogenate preparations from the O2k-Network: CZ Prague Krajcova A
- Krajčová A, Urban T, Megvinet D, Waldauf P, Balík M, Hlavička J, Budera P, Janoušek L, Pokorná E, Duška F (2020) High resolution respirometry to assess function of mitochondria in native homogenates of human heart muscle. PLOS ONE 15:e0226142. - »Bioblast link«
- Mitochondrial respiration is oxygen-dependent in human cardiac homogenate preparations from the O2k-Network: CZ Prague Krajcova A
2020 Week 4
- Jan-20 Mo
- IOC144 Innsbruck AT - O2k-Coaching Days fully booked
- Jan-21 Tu - IOC144 Innsbruck AT
- Mr Hai Wu (CEO of Huawei, Oroboros distributor in China) and Dr. Rupert Ecker (CEO of Vienna TissueGnostics) join the O2k-Coaching Days for extending collaborative projects.
- Jan-22 We - IOC144 Innsbruck AT
- Jan-23 Th - MitoFit Open Seminar - Erich Gnaiger: An improved index of obesity versus BMI - an issue of mitObesity and gender medicine in our modern world's obesogenic environment.
- Jan-24 Fr
- Further circulars to invited speakers for submission of tentative titles for oral and poster presentations - MitoEAGLE Summit Obergurgl 2020
- Do you write the serial number of the polarographic oxygen sensor (POS) in the ‘O2k configuration’ window?
- Including the POS serial number in the ´O2k configuration´ window is important to keep a track record of the sensors used according to the O2k. This will be especially helpful if troubleshooting is needed. Further details: O2k configuration
- Do you write the serial number of the polarographic oxygen sensor (POS) in the ‘O2k configuration’ window?
- Do you perform O2k-Chamber volume calibration?
- The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL, or the chosen volume. Further details: MiPNet22.11 O2k-FluoRespirometer manual
- Do you perform O2k-Chamber volume calibration?
- Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?
- The O2 solubility factor varies according to medium used and influence the air calibration. Further details: Oxygen solubility
- Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?